Creating and sharing knowledge for telecommunications

Precision of Sinefitting-Based Total Harmonic Distortion Estimator

Alegria, F.

Metrology and Measuring Systems Vol. XXIII, Nº 1, pp. 37 - 46, March, 2016.

ISSN (print): 0860-8229
ISSN (online):

Journal Impact Factor: 0,925 (in 2014)

Digital Object Identifier: 10.1515/mms-2016-0003

Download Full text PDF ( 1 MB)

Abstract
An analytical expression for the standard deviation of Total Harmonic Distortion (THD) estimation is derived. It applies to the case where the estimator uses sine fitting. It is shown that, in common circumstances, it is inversely proportional to the actual value of THD, the signal-to-noise ratio and the square root of the number of samples. The proposed expression is validated both with numerical simulations and an experimental setup using a Monte Carlo procedure.